A-107 Color Measurement Instruments
Sharp AR-C360P EFI Color Guide - 107
4. Plug the small connector on the adapter cable into the side of the
X-Rite DTP32 and plug the AC adapter into a wall outlet.
5. Turn on the computer.
6. Calibrate the DTP32 (see the following section).
7. Calibrate the Fiery using ColorWise Pro Tools and the DTP32 (see
page 2-61).
Calibrating the DTP32
You will need the black-and-white X-Rite Auto-Cal Strip included
with the densitometer. Calibrating the densitometer does not
require ColorWise Pro Tools.
To calibrate the X-Rite DTP32
1. Connect the densitometer to the computer and supply power (see
page A-105).
2. From the Main Menu on the X-Rite DTP32 display, press the p1
key once to reach p2.
3. Press the cal key.
Calibrating motor speed is displayed, followed by the words INSERT
CAL STRIP
.
AC adapter
Small connector