Sharp IQ-8000 Instrukcja Obsługi Strona 25

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 66
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 24
CarboVis 70x IQ TS Measuring / Operation
4 - 1
ba75950e02 11/2012
4 Measuring / Operation
4.1 Determination of measured values
Fig. 4-1 Determination of measured values
During a measurement, the absorbance spectrum of the test sample is
recorded and set against the stored absorbance spectrum of a zero
adjustment (measurement in ultrapure water). The algorithm for the
calculation depends on the selected application and/or the measuring
location. The resulting raw value can be adjusted to an independent
reference measurement with the aid of a so-called user calibration.
Absorbance spectrum
of zero adjustment
Absorbance spectrum
of measurement
Physical sensor, series number xxx
Main
sensor
software
Second. sensor
& virtual sensor
software
Main sensor meas. mode
(algorithm)
User calibration data
of main sensor
Main sensor raw value
Meas. value of main sensor
Secondary sensor
raw value& virtual sensor
Meas. value of sec. sensor
& virtual sensor
Second. sensor
& virtual sensor
meas. mode (algorithm)
User calibration data
of secondary sensor
& virtual sensor
IQ Sensor Net
Przeglądanie stron 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 65 66

Komentarze do niniejszej Instrukcji

Brak uwag