C35 Application Note

©2005-2013 SemiLEDs. All Right Reserved 
RoHS Compliant www.semileds.com
C35 Application Note V1.3
Subject to change without notice

Page
15
MeasurementandCalibration
IntegrationTime
Alongintegration timewillinducethermalissuesin LEDmeasurements. If theintegration timeis1
sec, the LED junction temperature (Tj) may reach over 50, which results in the light output dropping
more than 5%. It is recommended that the integration time is shorter than 25ms in high
power LED
measurements.
RecommendedMethod
1. UsemeasurementinstrumentswhichfollowtheCIE127standards.Theintegrationtimeshould
beshorterthan15ms.
2. Iftheoperatorusesnonstandardtesters,calibratethetesterwiththegoldensamplebefore
measurement.ThegoldensampleshouldbemeasuredbytheinstrumentfollowingCIE127
standards(ea.ISCAS
140B).